Applanation Tonometer Hans Heiss HAT-001 TYPE T

Sign-In/Register
to view price

HAT001 Type T Made in China

Based on the Goldmann

Principle which is a world widely accepted gold standard, T170 applanation tonometer can provide fast and accurate IOP measurement data to clinical diagnosis and treatment.

Category:
logo hans heiss

Description

HAT001 Type T Made in China

Based on the Goldmann

Principle which is a world widely accepted gold standard, T170 applanation tonometer can provide fast and accurate IOP measurement data to clinical diagnosis and treatment.

Adjustable mechanical

Design ensures the reliable measurement result and durable performance of the instrument. Repeated measuring procedures do not reduce the natural ocular pressure of the patient.

In combination with

All the tower type slit lamps. The measurement can be carried out on the patient sitting at the slit lamp in conjunction with other routine microscopic examinations.


Mount the Tonometer onto the Slit Lamp. Position the illumination of the Slit Lamp properly light up the Prism and switch the filter to the Cobalt Blue, in conjunction with the use of fluorescein, the 10P measurement value can be read from the dial drum of tomomter.


Specifications

Range of Measurement 0 mmHg ~ 80mmHg(0 kPa ~ 10.64kPa)
Moving Range of the Rings 1.53×2=3.06mm
Prism Diameter 7mm
Moving Range of Prism 0~3mm
Dimension 190mmx80mmx80mm
Net Weight 480g
Working Temperature 0°C~40°C / 32°F~104°F
Working Relative Humidity 10%~85%

Brand

Hans heiss

Hans Heiss
logo hans heiss